CaliX-L Level Transmitter Calibration

Precision calibration for radar, guided-wave radar and float-type level transmitters using controlled linear movement with 10-micron position resolution.

A Better Reference for Radar Verification

CaliX-L is Mezur's precision level-transmitter calibration service platform. Its 5-metre horizontal axis uses a servo-driven ball screw, encoder feedback, PLC control and touchscreen HMI to create known, repeatable movement with a position resolution of 10 microns.

For non-contact radar calibration, the transmitter is mounted on the moving platform and aimed at a fixed metallic target wall. As the platform moves toward or away from the wall, the system records the actual travel and the radar's measured distance or 4–20 mA output.

The two readings are compared point by point to quantify error, linearity and repeatability across the selected range. The CaliX-L reference motion is substantially finer than a typical radar accuracy of ±1 mm, enabling meaningful verification of transmitter performance.

CaliX-L five-metre linear level transmitter calibration platform with HMI and ball-screw drive
5 mHorizontal servo-controlled ball-screw axis
10 µmPosition resolution for precise movement
4–20 mATransmitter signal acquisition
3 TypesRadar, GWR and float transmitters

Known Movement. Measured Output. Direct Comparison.

The fixed target and moving transmitter create a controlled distance change that can be compared directly with the radar reading.

01

Mount & Connect

The radar is secured on the moving platform, aimed at the stationary metal wall and connected to the acquisition system.

02

Set the Reference

The platform is brought to its datum; servo, encoder and PLC establish the known starting position.

03

Move & Record

The ball screw moves the radar through manual or predefined points while actual position and radar output are logged.

04

Compare & Report

Radar readings are compared with actual travel to evaluate error, linearity and repeatability across the range.

Technical Specifications

Linear Mechanism5-metre precision ball screw
Drive & FeedbackServo motor with encoder
Position Resolution10 microns
Calibration Range0 to 15 metres, expandable by configuration
ControlPLC with touchscreen HMI
Operating ModesManual and semi-automatic
Signal Input4–20 mA analog
Supported InstrumentsRadar, GWR and float type
TraceabilityNABL-traceable calibration support

Accuracy Verification

The CaliX-L system has been calibrated by a NABL-accredited laboratory using a laser interferometer traceable to METAS, Switzerland.

Stated uncertainty

±(15 + 1.5L) µm, where L is in metres, at 95.45% confidence (k=2).

Reference Results

Set PointLaser ReadingError
0 mm0 mm0 µm
10 mm9.9862 mm13.8 µm
1,000 mm999.9236 mm76.4 µm
5,000 mm4,999.926 mm74 µm

Built for Level Measurement Users

Transmitter OEMs

Calibration & QA Labs

Chemical & Pharma

Oil & Gas

Food & Beverage

Maintenance Teams